Diagnostic monitor requirements and reporting workstations
Diagnostic monitor requirements under Polish law: monitors, resolution, DICOM GSDF calibration, lighting and testing for radiology reporting workstations.
A workstation for reporting digital radiology studies in Poland must meet the requirements of the Regulation of the Minister of Health of 11 January 2023 on the conditions for the safe use of ionising radiation for all types of medical exposure (Annex 1). For diagnostic monitor requirements, the regulation sets the number of monitors, their resolution and calibration, and the lighting in the room. Luminance and regular quality tests are covered by a separate regulation on operational tests. Legal status as of 25 November 2026.
In general radiology, a reporting workstation requires:
- at least two reporting monitors in portrait orientation working as a pair, with a pairing certificate from the manufacturer (or one monitor with two active areas)
- a resolution of at least 1024 × 1024 pixels
- calibration to the GSDF curve (Grayscale Standard Display Function) from DICOM, the standard format for medical images
- software with the functions listed in the regulation
- a room in which the illuminance on the monitor does not exceed 15 lux
Mammography has higher thresholds. Monitors undergo basic tests on every working day and specialist tests at least once every 12 months.
Legal basis
- Regulation of the Minister of Health of 11 January 2023 on the conditions for the safe use of ionising radiation for all types of medical exposure (Journal of Laws 2023, item 195; consolidated text Journal of Laws 2025, item 1288), text in ISAP (in Polish). Annex 1 sets out the requirements for reporting and review workstations.
- Regulation of the Minister of Health of 12 December 2022 on operational tests of radiological equipment and auxiliary equipment (Journal of Laws 2022, item 2759), text in ISAP (in Polish). Annex 2 sets out the monitor tests.
Reporting monitor and review monitor
The testing regulation distinguishes two types of monitor:
- a reporting monitor is used to present medical images so that a doctor can make diagnostic decisions
- a review monitor is used to present or view images for other purposes, for example in the referring doctor’s consulting room
Digital radiography studies, both computed radiography (CR) and direct digital radiography (DDR), must not be reported on a review monitor. Nor may they be reported from secondary images or printouts.
Every workstation, reporting or review, must use a computer with a dedicated graphics card that supports diagnostic monitors. The requirements do not apply to additional monitors used to operate the workstation, for example for the radiology information system (RIS) and email.
Requirements for a reporting workstation
| Requirement | General radiology | Mammography | CT, angiography |
|---|---|---|---|
| Monitors | at least 2 monochrome in portrait orientation, a pair with a pairing certificate, or 1 with two active areas | as in general radiology | at least 1; colour if images are stored in colour |
| Minimum resolution of the display area | 1024 × 1024 | 2048 × 2048 | 1024 × 1024 |
| Greyscale curve | DICOM GSDF | DICOM GSDF | DICOM GSDF |
| Illuminance on the monitor’s working surface | up to 15 lux | up to 10 lux | up to 15 lux |
| Room walls | dark, non-reflective | dark, non-reflective | dark, non-reflective |
Reporting workstation software in general radiology and mammography must allow at least:
- changing the display window
- splitting the screen into several images
- changing LUTs (look-up tables that remap pixel values)
- magnification of at least 4×
- linear calibration
- measuring distance and density in a selected region of interest (ROI)
- displaying a negative
- monitor calibration to the DICOM standard
For mammography screening, the software must show the full set of four images of the patient at the same time. In CT and angiography, DICOM software suited to the clinical scope of the images assessed is required.
Reporting workstations in dentistry follow the same requirements for monitor, resolution, GSDF curve and software as in CT. These requirements do not apply to intraoral images.
A digital radiology system must also provide lossless archiving protected against changes to the data.
Basic and specialist monitor tests
Test results prove that the workstation still meets the requirements. Before a test, the monitor must be switched on for the time recommended by the manufacturer or, if none is given, for at least 30 minutes. Measurements are taken under the lighting conditions of the reporting room.
Basic tests
These are performed on every day the monitor is used, before work begins. They check that:
- the screen surface is clean and free of scratches, with no reflections of light sources
- the standard test image shows no artefacts, faulty pixels or flicker, and its lines and greyscale fields can be told apart
- on workstations with two or more monitors, the test image looks identical, with the same brightness and tint
- threshold contrast is maintained, meaning the required number of letters and squares is visible in the test fields
Specialist tests
These are performed at least once every 12 months. Parameters are measured with instruments:
| Test | Reporting monitor | Review monitor |
|---|---|---|
| Luminance uniformity (centre and four corners) | max. 20% | max. 30% |
| Agreement of maximum luminance between the workstation’s monitors | max. 10% | max. 20% |
| Contrast (ratio of highest to lowest luminance) | min. 350 in mammography, min. 250 in general radiography, min. 100 in other cases | min. 100 |
| Deviation of the greyscale curve from DICOM GSDF | max. 10% | max. 20% |
Specialist tests are performed by an accredited laboratory or by a specialist in medical physics or medical engineering employed by the healthcare provider (Art. 33l(9) of the Atomic Law). Without such a specialist on staff, you order them from a laboratory accredited by the Polish Centre for Accreditation (PCA). After a repair that may affect image quality, the tests must be repeated. The repeat test must cover at least the parameters the repair may have affected (§ 4 of the testing regulation).
How we choose a reporting workstation
Workstations and diagnostic monitors are part of our services for radiology departments. We start with questions, not with a hardware catalogue:
- What the doctor reports: X-ray, mammography, CT, MRI. This determines the number and resolution of monitors and whether colour is needed.
- Which software they use. A viewer for the picture archiving and communication system (PACS) or the scanner manufacturer’s workstation has its own requirements for the processor, memory and graphics card, especially for 3D reconstructions.
- Where the workstation stands. The room must meet the wall and lighting requirements. In a new department, agree this with the contractor before the walls are painted.
- How the workstation connects to PACS: over a wired connection, in a separate network segment, with bandwidth matched to the size of the studies. In one project we expanded the network and cabling specifically for a reporting room.
- Who does testing and calibration: software for calibration and basic tests, and a schedule for specialist tests.
All the elements of a department, including RIS, PACS and disc publishers, are collected in our IT checklist for a new X-ray, CT or MRI department.
Teleradiology workstations
A doctor who reports studies remotely also works at a reporting workstation. The regulations make no exception for workstations outside the department. A monitor in the doctor’s home or in a reporting centre must therefore have the same parameters, calibration and test records. Our article on remote PACS access for reporting physicians shows how we set up the account, VPN and workstation for a new reporting physician.
Common mistakes
- An ordinary office monitor “with a good panel” used as a reporting monitor.
- Two monitors bought separately, without a pairing certificate.
- A reporting room with a window that cannot be darkened and bright ceiling lights.
- No records of the daily basic tests.
- Replacing a graphics card or monitor without checking whether the tests must be repeated.
Send us the list of studies you report, and we will match the workstations to them and to the requirements of the regulation.